FT-IR spectrometer, capable of meeting the needs of different requirements of research and development users. INOLI20 can optional configure automatic switching of MIR/NIR sources and detectors, it has the standard spectral resolution of better than 0.25cm-1 and expandability of spectral range from MIR to NIR (12800-50cm-1). INOLI20 can accommodate different internal and external accessories. It provides researchers another FTIR spectrometer to reach a new stage.
Superior Spectral resolution:
INOLI20 has spectral resolution of better than 0.25cm-1 in whole range and it can upgrade to higher spectral resolution if required. The superior resolution is suitable for most of R&D projects.
Expandable Spectral Range:
INOLI20 FTIR spectrometer can achieve the analysis of both MIR and NIR spectral range in one device, through switching the optical element automatically. It can expand from the standard MIR spectral range (8000-350cm-1) to NIR range(12800-4000cm-1). FT-NIR owns special advantages: under general circumstances, samples can be analyzed in vials directly without diluting, due to the fact that quartz is transparent in NIR range.
High Performance Design:
INOLI20 FTIR spectrometer designs exclusively for R&D institutions. It combines easy operations, high-performance design and high-cost performance into one device. INOLI20 configures room temperature DLaTGS detector and users can choose other high sensitive detectors, such as liquid nitrogen cooled MCT, Thermo Electric Cooler MCT,LN InSb and TEC InGaAs detectors. Those detectors can provide higher sensitivity and higher energy throughput for the experiment. All detectors have dual-channel A/D and 32 bit dynamic range, and acquire the lowest electronic noise.
TGA-IR coupling Module
GC-IR coupling Module
External sample compartment, vacuum or purgeable
External vacuum UHV Chamber
2cm to 20 meters length gas cells
Integrating sphere accessory
In-situ Transmission accessory
In-situ Diffuse Reflection accessory
ATR and specular accessory
Application
Pharm & Life Science
Protein conformation and quantification
Quantification for Active Pharmaceutical Ingredients and excipient in water solution
Microorganism Identification
Characterization for the volatility and stability of medicine combined with TGA module
Polymer and Chemical Products
Detection and Characterization for the volatility and decomposition combined with TGA module
Monitoring reaction process in lab combined with MIR fiber probe module
Surface analysis
Detection and Characterization for the Ultra-Thin film and mono-layer film
Characterization for the erosion process
Material Science
Detection of Emissivity of building materials
Evaluation for Optical material such as Infrared windows and mirrors
Study on black material with photoacoustic spectroscopy
Semiconductor Industry
Measurement for thickness of various non-metal films
Measurement for content of Carbon/Oxygen in silicon QC
Soil Analysis
Evaluation for nutrition and organic matter in soil
Study on the characteristic of different soil
|
Items |
Standard Parameters |
Optional Parameters |
||
|
Spectral Resolution |
< 0.3cm-1 |
|||
|
Spectral Range |
8000-350cm-1 |
12800-4000cm-1 |
||
|
Light Source |
Ceramic SiC |
tungsten lamp |
||
|
Beamsplitter |
KBr on Ge |
Quartz, CaF2, ZnSe, etc. |
||
|
Detector |
Room temperature DLaTGS |
LN MCT, TEC MCT,LN InSb, LN Ge and TEC InGaAs etc. |
||
|
Sample Compartment |
Transmission hold 50mm x 75mm |
Emission port, External compartment, External sealed UHV Chamber, etc. |
||
|
Signal to Noise |
≥40,000:1 |
|||
|
Photometric Precision |
≤0.1%T |
|||
|
Interferometer |
High stability Cube corner interferometer |
|||
|
Wavenumber Accuracy |
< 0.01 cm-1 |
|||
|
Wavenumber Precision |
< 0.1 cm-1 |
|||
|
Operating temperature |
-10°C to +40°C |
|||
|
Storage temperature |
-20°C to +55°C |
|||