IVD2004 Automatic Vacuum Distillation tester A special instrument for detecting the boiling point of partially or completely evaporated petroleum products. The instrument uses single chip microcomputer as the system control core, guided operation, the measurement process automation degree is high. Set the constant temperature of circulating cooling water and receiving chamber and the required decompression pressure according to the measurement needs, control the distillation speed and constant steam decompression pressure, cooling circulating water temperature and receiving chamber temperature. Comply with ASTM D1160
1.Smart heating management system to ensure that the distillation rate meets the requirements of the experimental method.
Record point user self-set:
The user can set the recovery volume to record the corresponding temperature
The user can set the temperature for recording the corresponding recovery volume
Automatic recording of specified recording points
Five ways to end the experiment:
(1) End of the end point: end the experiment when the final distillation point is detected.
(2) Dry point end: The experiment ends when the dry point is detected.
(3) End of temperature: End the experiment according to the temperature value set by the user, and print out.
(4) Volume end: End the experiment according to the volume value set by the user, and print out.
(5) Keyboard end: Press the exit key to end the experiment and print out the output.
2.Equipped with internal clock, no need to enter the experiment date.
| Atmospheric pressure temperature control range | Room temperature ~400℃ | ||||
| Resolution | 0.1 ℃ | ||||
| Temperature control accuracy | Plus or minus 0.3 ℃ | ||||
| Temperature detection | Distillation 0~1000W Continuous adjustable cold bath 500W | ||||
| Temperature control element | Special platinum resistance (Pt100) sensor | ||||
| Distillate rate | 2 ~9mL/min | ||||
| Condensation control temperature range | Room temperature ~90℃ arbitrary setting | ||||
| Accept room temperature control range | Room temperature ~80℃ arbitrary setting | ||||
| Reduced pressure control range | 130Pa ~ 6.7kPa Arbitrary setting | ||||
| Pressure accuracy | [kPa ±0.01] 1kPa | ||||
| Ambient temperature | - 10 ℃ ~ 45 ℃ | ||||
| Relative humidity | ≤ 70% | ||||
| power | 2200W | ||||
| Distillation power | 1000W | ||||
| Working power supply | AC220V 50/6OHz | ||||
| Overall dimension | 580mm×840mm×1010mm | ||||
| Instrument weight | 122kg | ||||